性能最高,尺寸最小的OEM模块,采用最新硬件平台,简单操作可与各种医疗设备系统集成。 1)在各种的临床应用环境中性能优异 2)具备心电门控抗运动功能 3)符合医疗设备第三版及Rohs要求 4) TTL通讯串口 5)小体积,扁平结构便于集成

技术规格
Physical Dimensions
| SPECIFICATIONS | VALUES | 
|---|---|
| Width | 1.0 in [25.4 mm] | 
| Length | 2.75 in [69.9 mm] | 
| Height | 0.72 in [18.3 mm] | 
| Weight | 1.25 oz. [35.4 g] | 
Measurement
| SPECIFICATIONS | VALUES | 
|---|---|
| Technique | Oscillometric | 
| NIBP Accuracy | Meets ANSI/AAMI SP10, EN 1060-4, ISO 81060-2 | 
| Patient Applications | Adult, Pediatric, Neonatal | 
| Systolic Range | 
					
  | 
			
| Diastolic Range | 
					
  | 
			
| Pulse Rate Range | 
					
  | 
			
| Pulse Rate Accuracy | ±2% or ±3 BPM, whichever is greater | 
| Transducer Accuracy | ± 3 mmHg over full range in operating conditions | 
| Operating Modes | Manual | 
| Operating Conditions | 0˚C to 55˚C, 15% to 90%, non-condensing humidity | 
| Storage Conditions | -20˚C to +65˚C, up to 90% RH non-condensing | 
Electrical
| SPECIFICATIONS | VALUES | 
|---|---|
| Operating Voltage | 5 VDC ± 5%, 6 VDC ± 5% | 
| Power Consumption | 
					Using 5 VDC Nominal
					
  | 
			
| Communication Protocols | TTL (5V) | 
| Safety & Regulatory Standards | IEC 60601-1, IEC/EN80601-2-30, EN60601-1-2:2007, ISO 81060-2:2009 | 
Advantage Technologies
| SPECIFICATIONS | VALUES | 
|---|---|
| RMT-MAX (Maximum Performance R-wave Motion Tolerance) | Yes | 
| TMT-MAX (Maximum Performance Transport Motion Tolerance) | Yes | 
| VET-S (Standard Veterinary Monitoring) | Yes | 
Customer Support
| SPECIFICATIONS | VALUES | 
|---|---|
| Standard Support | Yes | 
| Custom Support | 10 hours* | 
| Custom Design | Additional Charge | 
Warranty
| SPECIFICATIONS | VALUES | 
|---|---|
| Time | 1 year | 
| Cycles | 30000 | 
						